Compact absorbance microplate readers for ELISA
The systems’ innovative design combines eight measurement channels with long-life LED lamps for fast parallel reading and minimal maintenance costs.
The LT-4500 is Tecan’s reliable, validated and compact absorbance microplate reader for ELISA. The LT-4500 provide accurate, fast photometric measurements using innovative eight-channel optics and LED technology.
Specifications
Detection mode | Absorbance | |
Light source | LED, self-calibrating | |
Photodetector | Silicon photodiodes | |
Plate type | 96-well microplates (including strip-well microplates) | |
Shaking | Linear shaking in four different modes | |
Wavelength range | 400-750 nm | |
Wavelength selection | Up to eight filters on wheel | |
Measurement modes | Endpoint, kinetic and multilabel | |
Measurement channels | 8 | |
Reference channel | 1 | |
Measurement range | 0-4 OD | |
Resolution | 0.0001 OD | |
Precision @450/492nm | 0.000-2.000 OD ≤(0.5 % + 0.005 OD) | |
Precision @450/492nm | 2.000-3.000 OD ≤(1 % + 0.005 OD) | |
Accuracy @450/492nm | 0.000-2.000 OD ≤(0.5 % + 0.010 OD) typical* | |
Accuracy @450/492nm | 2.000-3.000 OD ≤(1 % + 0.010 OD) typical* |
* Specifications are subject to change. Typical performance values represent the average observed factory tested values.
Highlights
- Ideal for absorbance-based ELISAs, kinetic and multi-label reading
- Ready for qualitative and quantitative data analysis using Magellan data analysis software
- All major curve fittings supported
- Data can be exported to the LIS (Laboratory Information System)
- Magellan’s example files and tutorials enable you to get started easily
- 405, 450, 492 and 620 nm filters included
- Eight-channel optics for fast reading
- Reference channel for optimized light intensity
- Long-life LED lamps offer 10 times longer maintenance-free operation than conventional halogen lamps
- Lamp changes are a thing of the past, helping to save money
- Two year warranty and tested reliability of >100,000 plates
- Linear plate shaking
- Designed to meet IVDD 98/79/EC and RoHS directive 2011/65/EU
- Provides FDA 21 CFR Part 11 functionality for regulated environments via Magellan Tracker software
- MultiCheck™ QC plate for IQOQ