Near Infrared Spectrometer
The Apo-Ident analysis system, a ready-to-use solution for pharmacy laboratories. The integrated, validated substance database (chemometric model) enables identity verification within seconds instead of long, time-consuming wet chemical analyses.
For its manufacturing processes, the food industry is dependent on monitoring ingredients such as fat, protein and water. Critical process parameters are recorded quickly and permanently using NIR spectrometers, examine the production process to the second for agricultural, food and meat samples. With the implementation of the Finder SD Rotator analyzer in the production process, better food can be produced more efficiently by the time-saving and non-destructive analysis of solid samples. Due to the Scanning Grating Technology that is used, the system is fast, robust and reliable. The creation of an ensured chemometric model that meets all technical and regulatory requirements is the most important factor for using NIR in the food industry.
The dust-proof and water-resistant Finder SD Rotator analyzer can be integrated at any workplace and for a variety of functions. The rotator for samples e.g. in a petri dish provides reproducible results. Due to the instrument‘s calibration standards the system constantly checks itself or is recalibrated. A fast and easy integration into user-specific applications is possible with HiperScan instrument drivers.
Benefits for your industry
• Quality assessment by measuring content of fat, protein and water
• Automatic detection of concentrations
• Analysis of dairy and meat products, baking ingredients and grains
Advantage overview
• Robust design, IP65 certified
• Patented Fraunhofer MEMS scanning grating technology
• Thermally stabilised
• Integrated stabilised Wolfram halogen light source
• Easy to clean, no cross-contamination
• Rotating sample dish Ø 94mm
• Analysis results in a few seconds
• User-friendly software
Specifications
Spectral range | 1,000–1,900 nm |
---|---|
Spectral resolution | 10 nm* |
Stray light | ‹ 0.2 % |
Measuring time | ‹ 5 s (averaging over 500 scans), configurable 1–15 s |
Detector | InGaAs single detector, uncooled |
Wavelength accuracy | ± 0.5 nm* |
Wavelength reproducibility | ± 0.2 nm* |
Photometric reproducibility | ± 0.1 %* |
SNR | > 2,000:1 (averaging over 2,000 scans)* | > 1,000:1 (averaging over 500 scans)* |
Photometric linearity deviation (max/RMS) | 1 % / < 1 %* |
Light source | Tungsten halogen lamp |
Probe/optical input | Diffuse reflection, 23 mm Ø; petri dish 94 mm |
Thermal stabilization | Yes |
Dimensions | 225mm x 271mm x 460mm |
Operating temperature | Typ 55: 15 to 35 °C or Typ 50: 5 to 30 ° C |
Storage temperature | -20 to 60 °C (non-condensing) |